New TopShim run:

version = topshim 2.1.0
date = 2021-12-22
time = 11:55:48.420
command = topshim tunea

FINE SHIMMING

Parameters:

probe = Z168276_0016
solvent = H2O+DMSO
shim nucleus = 1H
nucleus optimised for = 1H
o1p (set) = 4.70 ppm

Results:


1D Shimming:
Iteration 1: FWHM Before= 2.6 Hz, Target= 0.4 Hz, After= 0.5 Hz
Iteration 2: FWHM Before= 0.5 Hz, Target= 0.4 Hz, After= 0.5 Hz
Iteration 3: FWHM Before= 0.5 Hz, Target= 0.4 Hz, After= 0.6 Hz
Iteration 4: FWHM Before= 0.6 Hz, Target= 0.4 Hz, After= 0.3 Hz
Iteration 5: FWHM Before= 0.3 Hz, Target= 0.4 Hz, After= 0.6 Hz
Iteration 6: FWHM Before= 0.6 Hz, Target= 0.4 Hz, After= 0.4 Hz
Shim succeeded

Lineshape quality factor (1D spectrum, spinning) = 0.658


Performing tune after shimming
tune lock level increase = 1.0

shim changes: 
Z	+14
Z2	+25
Z3	+53
Z4	-31
Z5	-275
Z6	-690
Z7	+611
Z8	+7867
X	+3
XZ	+8
Y	+39

duration = 1 min 41 sec
completed successfully
finished Wed Dec 22 11:57:28 2021


========================================================================