New TopShim run:

version = topshim 2.1.0
date = 2021-12-22
time = 13:23:43.728
command = topshim tunea

FINE SHIMMING

Parameters:

probe = Z168276_0016
solvent = H2O+DMSO
shim nucleus = 1H
nucleus optimised for = 1H
o1p (set) = 4.70 ppm

Results:


1D Shimming:
Iteration 1: FWHM Before= 0.8 Hz, Target= 0.4 Hz, After= 2.3 Hz
Iteration 2: FWHM Before= 2.3 Hz, Target= 0.3 Hz, After= 0.5 Hz
Iteration 3: FWHM Before= 0.5 Hz, Target= 0.3 Hz, After= 1.2 Hz
Iteration 4: FWHM Before= 1.2 Hz, Target= 0.4 Hz, After= 0.6 Hz
Iteration 5: FWHM Before= 0.6 Hz, Target= 0.4 Hz, After= 1.2 Hz
Iteration 6: FWHM Before= 1.2 Hz, Target= 0.4 Hz, After= 0.3 Hz
Shim succeeded

Lineshape quality factor (1D spectrum, spinning) = 0.499


Performing tune after shimming
tune lock level increase = 1.0

shim changes: 
Z	+118
Z2	+15
Z3	-14
Z4	-83
Z5	-24
Z6	-480
Z7	-420
Z8	+7268
X	+11
XZ	+135
Y	+1
YZ	+39

duration = 1 min 45 sec
completed successfully