
New TopShim run:

version = topshim 2.1.0
date = 2021-12-23
time = 11:30:39.887
command = topshim tunea

FINE SHIMMING

Parameters:

probe = Z168276_0016
solvent = H2O+DMSO
shim nucleus = 1H
nucleus optimised for = 1H
o1p (set) = 4.70 ppm

Results:


1D Shimming:
Iteration 1: FWHM Before= 4.8 Hz, Target= 0.4 Hz, After= 1.0 Hz
Iteration 2: FWHM Before= 1.0 Hz, Target= 0.4 Hz, After= 0.7 Hz
Iteration 3: FWHM Before= 0.7 Hz, Target= 0.4 Hz, After= 0.3 Hz
Iteration 4: FWHM Before= 0.3 Hz, Target= 0.4 Hz, After= 0.5 Hz
Iteration 5: FWHM Before= 0.5 Hz, Target= 0.4 Hz, After= 0.4 Hz
Shim succeeded

Lineshape quality factor (1D spectrum, spinning) = 0.468


Performing tune after shimming
tune lock level increase = 1.0

shim changes: 
Z	+101
Z2	+22
Z3	+38
Z4	-175
Z5	-134
Z6	-72
Z7	-147
Z8	+6205
X	+5
XZ	+28
Y	+62
YZ	+33

duration = 1 min 44 sec
completed successfully
finished Thu Dec 23 11:32:22 2021


========================================================================
