New TopShim run:

version = topshim 2.0.0
date = 2023-10-19
time = 10:45:52.584
command = topshim  z6 1d optsolvent coil=1H tuneboff tuneaoff

FINE SHIMMING

Parameters:

probe = Z161159_0002
solvent = H2O+D2O_salt
shim nucleus = 1H
nucleus optimised for = 1H
o1p (from lock) = 4.70 ppm

Results:


1D Shimming:
Iteration  1: FWHM Before= 3.9 Hz, Target= 0.3 Hz, After= 0.4 Hz
Iteration  2: FWHM Before= 0.4 Hz, Target= 0.1 Hz, After= 0.6 Hz
Iteration  3: FWHM Before= 0.6 Hz, Target= 0.1 Hz, After= 0.2 Hz
Iteration  4: FWHM Before= 0.2 Hz, Target= 0.1 Hz, After= 0.3 Hz
Iteration  5: FWHM Before= 0.3 Hz, Target= 0.1 Hz, After= 0.2 Hz
Iteration  6: FWHM Before= 0.2 Hz, Target= 0.1 Hz, After= 0.4 Hz
Iteration  7: FWHM Before= 0.4 Hz, Target= 0.1 Hz, After= 0.3 Hz
Iteration  8: FWHM Before= 0.3 Hz, Target= 0.1 Hz, After= 0.2 Hz
Slow convergence detected. Shimming was aborted.
Using the best result as the final shim state.

Lineshape quality factor (1D spectrum, spinning) = 0.291

shim changes: 
Z	-515
Z2	-249
Z3	-716
Z4	-2314
Z5	-415
Z6	+4079

duration = 53 sec
completed successfully
finished Thu Oct 19 10:46:44 2023


========================================================================